ALLURIS TMI Serie Instrucciones De Servicio página 21

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Overload
Modes of operation
Display type
Update rate
(Standard)
Update rate (Peak)
Thresholds mode of
operation
Data memory
Energy supply
Battery durability
Low Batt indicator
Interface
Temperature range
(operation)
Temperature range
(storage)
Weight
Size (L x B x H)
Casing material
Type of protection
The instrument, apart from expendable and wearing parts, is maintenance-free. Depending on the area and
frequency of use, torque meters should be regularly, namely at least once a year, be checked and calibrated. Our
calibration service (e-mail address: [email protected]) includes the instrument`s technical check, calibration –
factory calibration – and, if necessary, adjustment of the measurement range.
12.0 Troubleshooting
Display fades out!
Check the remaining voltage. Plug in the adaptor/charger and check the electricity connection.
Values are depicted with inverted prefix!
Change the mode of operation or the prefix; see Chapter 6.4.
No deletion of selected values!
The instrument deletes stored values sequentially; the lastly measured value is deleted first. Selective
deletion is not possible.
Threshold monitoring, data storage and other functions do not work!
Check your instrument`s model. The TMI-100 model is not equipped with the same functions as the TMI-
200 model.
RS-232C communication is disrupted!
Please check the connection cable`s wiring as well as the feed line`s compliance with the PC-card`s
input assignments.
TMI-300; printing is not possible!
200% (F.S.) | warning at 120% of the measuring range
Standard, right-handed, left-handed
(showing peak-values as Nm, Ncm, Kgcm or Lb-in)
LCD, 4-digits, 12mm high with side indication 7mm
125 msec | 250 msec | 500 msec | 1000 msec (adjustable)
Internal, rechargeable NiMH battery | Universal power supply- / battery charger
8,0 Kg
12,5 Kg
Aluminium cast, protection cover PS
Threshold monitoring (adjustable thresholds and
positive/negative indication by coloured LEDs)
1 msec
quantity : 1000 measurements; statistics function: single,
average, max, min
100...240VDC (50Hz)
8...12 h
yes
0°...40° (max. 85%rF)
-10°...50 (dry storage)
8,0 Kg
320 x 227 x 126 mm
IP44 (splash-water protection)
19
6-colours LEDs
USB 1.1
Hirose socket
12,5 Kg
8,0 Kg
12,5 Kg
Tabla de contenido
loading

Este manual también es adecuado para:

Tmi-100Tmi-200Tmi-300

Tabla de contenido