Table 2. Product specifications (continued)
Effective isotropic
radiated power (EIRP)
Operating range
Transmitter expected
service life
Transmitter wireless communication
Quality of service
The transmitter and insulin pump connect via smart device connectivity. The transmitter sends
glucose data and system-related alerts to the pump. The pump verifies the integrity of received data
after wireless transmission.
Data security
The transmitter is designed to only accept radio frequency (RF) communications from recognized and
linked devices. You must pair your pump with the transmitter before the pump will accept information
from the transmitter.
The MiniMed insulin pumps and system components (meters and transmitters) ensure data security
via proprietary means and data integrity using error checking processes, such as cyclic redundancy
checks.
Traveling by air
Your transmitter is safe for use on commercial airlines. If questioned by airline personnel about the use
of your device, please show them your Medical emergency card.
-9.9 dBm (0.1 mW)
Up to 1.8 meters (6 feet) in free-air
The transmitter expected service life is one year depending on patient
usage.
38