Operating Instructions
1. Locate and position the FreeStyle
location for use.
a. The concentrator should be well-ventilated so that
the air intake filters and air outlet are not obstructed.
b. The concentrator should be positioned so that all
audible and visual indicators and alarms can be
easily seen and heard.
c. Be sure the air intake filters are in place before
operating the unit. If the unit is missing a filter,
contact your equipment provider.
2. Install a Battery Pack in the concentrator.
a. Align the battery with the bottom housing from the
center of the FreeStyle
b. Slide the battery inward until you hear a click to
confirm that it has locked in place.
®
Comfort
®
in a proper
®
Comfort
®
.
FreeStyle Comfort
3. Insert the AC or DC power supply into the power port
receptacle on the front, bottom right side of the unit.
4. Connect a nasal cannula to the oxygen outlet barb.
Ensure the connection is secure before operation.
NOTE: Any cannula(s) used other than the recommended
may affect performance.
Note: To Equipment Provider: The following oxygen
administration accessories are recommended for
use with the FreeStyle Comfort:
• Nasal Cannula with 7 feet (2.1 m) of tubing (Max-
imum LPM: 6 LPM): CAIRE Part Number 5408-SEQ
• Firebreak: CAIRE Part Number 20629671
A firebreak is recommended for use with any
cannula.
• CAIRE offers a firebreak intended to be used
in conjunction with the oxygen concentrator. The
firebreak is a thermal fuse to stop the flow of gas in
the event that the downstream cannula or oxygen
tubing is ignited and burns to the firebreak. It is
placed in-line with the nasal cannula or oxygen
tubing between the patient and the oxygen outlet
of the FreeStyle Comfort. For proper use of the
firebreak, always refer to the manufacturer's
instructions (included with each firebreak kit).
• For any additional recommended accessories,
please see the Accessories Catalog (PN ML-
LOX0010) available on www.caireinc.com.
PN MN250-C4 H | User Manual