4.3 Technical specifications

Name and model
Display system
Power supply
Measuring range
Accuracy of
measurement
(laboratory readings)
Clinical repeatability : 0.18 °C
Duration of
measurement
Memory capacity
Fever alarm
Automatic switch-off : approx. 3 minutes

Operating

temperature
Storage/transport
temperature
Dimensions
Weight
Article No.
EAN Code
The CE-mark of this device refers to the EC-directive 93/42/EEC
MDD.
The unit is designed for home use only.
For commercial use, a metrological inspection must be carried
out once every year. This calibration will be charged for and can
be carried out by an appropriate authority or an authorised
service centre - in accordance with the regulations for users of
medical equipment.
Electromagnetic compatibility:
The device complies with the EN 60601-1-2 standard for elec-
tromagnetic compatibility. Inquire at MEDISANA for details on
this measurement data.
In accordance with our policy of continual product
improvement, we reserve the right to make technical
and optical changes without notice.
The current version of this instruction manual can be found under
www.medisana.com

4 Miscellaneous

: MEDISANA Infrared thermometer FTN
: digital display (LCD)
: 3 V= , 2 batteries 1.5 V
(type LR03, AAA)
:
Body mode 34 °C to 42.2 °C
Object mode 0 °C to 100 °C
: in the region of
34 °C to 35.9 °C: ±0.3 °C
36 °C to 39 °C: ±0.2 °C
39.1 °C to 42.2 °C: ±0.3 °C
(under operating conditions)
: approx. 1 second
:
up to 30 readings
: for temperatures >37.5 °C
after the last measurement
Body mode 16 °C to 40 °C
:
Object mode 5 °C to 40 °C
: -20 °C to +50 °C
: approx. 147 x 38 x 21 mm
: approx. 48 g including batteries
: 76120
: 40 15588 76120 1
GB
27

Hide quick links:

Tabla de contenido
loading

Este manual también es adecuado para:

76120

Tabla de contenido