(2)Common installation errors
Bubble
1
If the bubbles on the surface of the medium are large and the bubble layer is thick,
measurement errors will be caused and even reflected ultrasonic waves will not be
received. Please take measures to prevent bubble generation, or install the
sensor in the bypass pipe for measurement. Other measuring instruments can
also be used, such as radar level gauge or magnetostrictive level gauge.
The sensor installation direction is wrong
2
If the sensor is not installed on the surface of the medium, the measurement
signal will be weakened. In order to ensure the best measurement effect, please
align the axis of the sensor with the surface of the medium, that is, perpendicular
to the surface of the measured interface.
bubble
Location with bubbles