3.6.1. The lower LCD display will have a default setting for the dose of the drug that you have just selected. If
the dose is correct, press the VOLUME key. If the dose is incorrect, enter the correct dose by using the
numeric keypad followed by the VOLUME key.
Note: The default units of grams or milligrams cannot be changed.
3.7. ENTER THE CONCENTRATION:
3.7.1. The lower LCD display will have a default setting for the concentration of the drug that you have just
selected. If the concentration is correct, press the VOLUME key. If the concentration is incorrect, enter
the correct concentration by using the numeric keypad followed by the VOLUME key.
3.8. ENTER THE DATE:
3.8.1. Enter the date by using the numeric keypad followed by the VOLUME key. You can enter a slash to
separate the numbers by using the DECIMAL POINT key on the numeric keypad.
3.9. ENTER THE LOT NUMBER:
3.9.1. Enter up to a fourteen digit lot number by using the numeric keypad followed by the VOLUME key. Press
the DECIMAL POINT key on the keypad to enter a "-" in the lot number.
3.10. ENTER THE BARCODE:
3.10.1. Enter up to a fifteen digit barcode by using the numeric keypad followed by the VOLUME key. Do not
place a "-" in the barcode by pressing the DECIMAL POINT key on the numeric keypad.
3.11. ENTER THE NUMBER OF LABELS TO PRINT:
3.11.1.
Enter the number of labels to print by using the numeric keypad followed by the VOLUME key.
3.12. START THE PRINTING PROCESS:
3.12.1. Press the START/STOP key on the keypad.
NOTES:
A. Each press of the STORE MEMORY key will return to the previous prompt in Label Mode.
B. The PUMP/DILUTE key will escape from Label Mode back to the last set-up of the pump prior to entering
Label Mode. This will maintain the calibration and other programmed settings of the pump.
The Reset Button or the On/Off button of the pump will both put the pump into the Initial default settings, which
occur when the pump is originally turned on. IF YOU SHUT OFF POWER OR RESET THE PUMP, THE
CALIBRATION AND OTHER PROGRAMMED SETTINGS OF THE PUMP WILL BE RESET TO THE
DEFAULT SETTINGS.
C. The label printing may be stopped by turning off the printer.
4.
Correcting Label Jams in the Printer
CAUTION: TURN POWER OFF TO THE PRINTER BEFORE MAINTENANCE AND DO NOT UNDER ANY
CIRCUMSTANCE, PLACE ANY SHARP, HARD OR ABRASIVE OBJECT SUCH AS A KNIFE OR SCREWDRIVER
INTO THE PRINTER TO REMOVE A LABEL!!
4.1. Open the printer cover which is located on top of the printer by sliding green Cover Release forward on each
side of the printer to unlatch it and then tilt it upward to open.
4.2. Carefully remove the jammed/stuck label(s) if possible and the roll of labels if needed.
4.3. If needed follow the procedure for cleaning the thermal print head in LP2824 Thermal Printer Quick Start.
4.4. If needed follow the procedure for cleaning the Platen in LP2824 Thermal Printer Quick Start.
CAUTION: DO NOT USE ANY ABRASIVE MATERIAL; IT MAY DAMAGE THE PRINTING MECHANISM!!
5.
Custom Label Procedure
5.1. ENTER CUSTOM LABEL MODE:
5.1.1. With the Pharm-Assist in the pumping mode (not the label mode) enter 1, 0, 1 followed by RECALL
MEMORY.
5.2. SELECT A LABEL MEMORY LOCATION:
-10-