Mectron Piezosurgery touch Manual De Uso página 20

Ocultar thumbs Ver también para Piezosurgery touch:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 142
IRRIGATION
The capacity of the peristaltic pump can be
adjusted by selecting the numbers on the
touch keyboard in the "irrigation" column.
7 capacity levels are foreseen:
0 = operation of the pump is closed: no
EN
irrigation outflows from the insert
From 1 to 6= the pump flow goes from
8 ml/min to approximately 75 ml/min.
The possibility of choosing the irrigation
capacity levels is tied to the type of function
selected, in the following manner:
ENDO - 7 flow levels: from 0 to 6
(from 0 to approximately 75ml/min)
PERIO - 7 flow levels: from 0 to 6
(from 0 to approximately 75ml/min)
SPECIAL - 6 flow levels: from 1 to 6
(from 8 to approximately 75ml/min)
CANCELLOUS - 6 flow levels: from 1 to 6
(from 8 to approximately 75ml/min)
CORTICAL - 6 flow levels: from 1 to 6
(from 8 to approximately 75ml/min)
IMPLANT - 6 flow levels: from 1 to 6
(from 8 to approximately 75ml/min)
PLEASE NOTE treatment without irrigation
is possible only with the ENDO and PERIO
functions, setting the irrigation capacity level
on "0".
In the event that an irrigation capacity
lower than 8 ml/min is required, use the
"bone grafting kit" (accessory that can be
ordered separately), inserting it between
the flow adjuster and the tube in silicon of
the handpiece, making it pass through the
peristaltic pump, and selecting 1 as the level of
irrigation.
WARNING: If the "bone grafting
kit" tube is kept inserted for the entire
duration of the intervention, the capacity
of the pump on all its levels is limited,
independently of which insert is used.
16
Q
(ref.
inside cover)
irrigation
function
6
implant
5
cortical
4
cancellous
3
special
2
perio
1
endo
0
OPTIONAL
light
auto
on
off
pump/clean