6.
Press the Start button.
The LOW CURRENT message
will turn up:
7.
Turning the Test current
control clockwise,
increase current until
obtaining the desired
value or until the current
bargraph reads 100
divisions. The bargraph
will show the test current
value as the percentage of
the nominal value selected
by using the Range
selector.
8.
The equipment should be connected to the mains supply in order
to work with 200 A.
9.
The lowest current for measuring is 10% of the range nominal
value. It is important to consider that the measurement errors
increase while test current decreases. The error is specified for
test currents higher than 80%.
10. The Display will show the resistance value measured and the unit
used (Ω [ohms], mΩ [milli-ohms] or µΩ [micro-ohms]).
11. The value can be retained in
the second line of the display
by pressing the button Hold. If
you press this button again,
the value will be erased.