TROUBLESHOOTING
If any abnormality arises during use, please check the following points
SYMPTOMS
Unit does not turn
Batteries have
on when the START/
run down.
STOP button is
pushed.
Battery polarities
have been positioned
incorrectly.
EE measurement error
Cuff has been placed
symbol shown on
incorrectly.
display or the blood
pressure value is
Did you talk or move
displayed excessively
during measurement?
low (or high).
Shaking of the arm with
the cuff on.
E1 error symbol shown
Air circuit abnormality.
on display
Cuff tube may not be
plugged into monitor
correctly.
E2 error symbol shown
Inflation pressure
on display
exceeding 300 mmHg.
E3 error symbol shown
Error determining
on display
measurement data.
EP error symbol shown
System error.
on display
Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no
circumstance should you disassemble or attempt to repair the unit by yourself.
Contact information for HoMedics Consumer Relations Department can be found
on the warranty page.
POSSIBLE
CORRECTION
CAUSES
Replace them with
four new "AAA"
alkaline batteries.
Re-insert the batteries in
the correct positions.
Wrap the cuff properly so
that it is positioned
correctly. Measure again.
Keep arm steady during
measurement. Measure
again.
Check cuff connection.
Measure again.
Switch the unit off, then
measure again.
Measure again.
Measure again.
28
SPECIFICATIONS
Power Source:
Four 1.5V DC "AA" alkaline batteries (not included)
Measurement Method: Oscillometric
Pressure: 40~280 millimeters mercury (mmHg)
Measurement Range:
Pulse: 40~199 beats/minute
Pressure: ±3 mmHg
Accuracy:
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Pump driven
Deflation:
Automatic Pressure release valve
Memory Capacity:
99 memories
Auto-shut-off:
1 minute after last button operation
Operation
Temperature: 41°F~104°F (5°C ~40°C)
Environment:
Humidity: 15% ~ 93% RH
Storage/Transportation
Temperature: -13°F~158°F (-25°C~70°C)
Humidity: Less Than 93% RH
Environment:
Weight:
0.63 lbs (286 g) (without batteries)
Standard size cuff: 9"–13" (23-33 cm)
Arm Circumference:
Large size cuff (not included):13"–17" (33-43 cm)
4.1"(L) x 5.7"(W) x 2.8"(H)
Dimensions:
105mm(L) x 145mm(W) x 70mm(H)
Accessories:
Arm cuff with tube
Internally Powered equipment
Type BF Equipment
IPx0-No special protection against the
ingress of water
Not suitable for use in presence of flammable
anesthetic mixture with air, oxygen or
nitrous oxide.
Continuous operation with short-time loading.
Follow instructions for use.
Note: These specifications are subject to change without notice.