After pressing Start with the desired reference type selected - example,
1-Jumper|Normal, the reference calibration process will begin. Once the
reference is set the user interface will indicate which end of the jumper to
disconnect from the module, and whether additional jumpers need to be
EN
attached before testing.
The Calibration Reference
results will be displayed on
a range from Good-to-Poor.
Begin testing only when a
Good quality reference is
achieved. If Average or Poor
is shown, follow the on-
screen recommendations
to improve performance.
Clean the connectors of
the reference cords and
the FiberTEK IV modules,
replace worn/damaged
cords.
Always use fibre optic
specific cleaning products
and 99% isopropanol/IPA,
never use rubbing alcohol.
Use extreme care when
cleaning module ports to
prevent damage.