Calibration Checks - Testing Instructions; (For Trained Specialist Staff Only) - boso medicus family 4 Manual Del Usuario

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Calibration checks - Testing instructions

(for trained specialist staff only)

Calibration checks – every 2 years at the
latest – must be performed by the follow-
ing institutions or persons:
– Manufacturer
– Trade measurements authorities
– Persons who fulfil the requirements of
the legislation governing the operation
of medical devices.
A) Function testing
Function testing of the monitor can
only be carried out on a person or with
a suitable simulator.
B) Testing of pressure circuit integrity
and deviation of pressure display
72
Note:
a) When the pressure in measurement
mode increases above 299 mmHg, the
quick release valve is triggered and the
pressure circuit opens. When the press-
ure in test mode increases above 320
mmHg, the measured value display
flashes.
b) For the calibration checks a special test
plug must be used. The test plug is
available for a fee from the manufac-
turer.
A pump ball must also be introduced
into the pressure circuit.
Tabla de contenido
loading

Tabla de contenido