If the measured field strength in the location in which the PinkWave™ is used exceeds
the applicable RF compliance level above, the PinkWave™ should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the PinkWave™.
b - Over the frequency range 150 kHz to 80 MHz, field strength should be less than 10 V/m.
Recommended Separation Distances Between Portable and Mobile RF
Communications Equipment and the "PinkWave™ Curing Light"
The PinkWave™ is intended for use in the electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the PinkWave™ can help
prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the PinkWave™
Curing Light as recommended below, according to the maximum output power of the
communication equipment. In general, portable RF communications equipment may
degrade performance if used closer than 30 cm (12 inches).
Mobile RF Communications Equipment and the "PinkWave™ Curing Light"
Rated Maximum
Output Power of
Transmitter (W)
0.01
0.1
1
10
100
Table: According to IEC 60601-1-2, Edition 4.0
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be determined using the equation applicable
to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE: At 80 MHz and 800 MHz, the separation distance for the
higher frequency range applies.
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Recommended Separation Distances Between Portable and
Separation Distance According To Frequency of Transmitter (m)
150 kHz to 80 MHz
outside ISM bands
80 MHz in ISM
3.5
d = [
] P
V
1
0.12
0.37
1.17
3.7
11.7
150 kHz to
80 MHz to
800 MHz
12
d = [
] P
d = [
V
2
0.2
0.64
2
6.4
20
800 MHz
to 2.5 GHz
12
23
] P
d = [
E
E
1
1
0.4
1
1.3
2.6
4
8
13
26
40
80
] P