TROUBLESHOOTING
To check the Self-Diagnostic Warning Light, turn on the key and count the number of flashes
Flash
Description
1
Low Battery
Low Battery Fault Battery charge is empty.
2
3
Battery
too high
4
Current limit
time-out
5
Brake fault
6
Out of neutral at
power-up
7
Speed Pot Fault
8
Motor Voltage
Fault
9
Other error
The batteries are running low.
•Recharge the batteries
• Recharge the batteries.
voltage
Battery voltage is too high. This may occur if overcharged &/or traveling
down a long slope.
• Turn the scooter power off then on again
• If traveling down a slope, reduce your speed.
The motor has been exceeding its maximum current rating for too long.
This may be due to a faulty motor
• Check the motor and associated connections and wiring.
• Turn the controller off, leave for a few minutes and turn back on again.
Either a park brake release switch is active or the park brake is faulty.
• Check if the lever is in Automatic Mode (D).
• If this flash code does not appear until the throttle has been moved out
of neutral, check for a short in the park-brake circuit.
• If this flash code appears at power-up, check for an open circuit in the
park-brake circuit.
• Check the park brake and associated connections and wiring
The Throttle is out of neutral when turning the controller on.
• Ensure the throttle is in neutral when turning the scooter on.
• The Throttle may require re-calibration.
• Check the its associated connections and wiring.
The throttle, speed limit pot or their associated wiring may be faulty.
• Check the throttle and speed pot and associated connections and wiring.
The motor or its associated wiring is faulty.
• Check the motor and associated connections and wiring.
The controller may have an internal fault.
• Check all connections & wiring.