H i k o n e b | 9 0 8 D C
Us e r Ma n u a l
Examples of surface disinfectants:
- Incidin® Plus *
- Incidin® Perfect *
- Antiseptica Kombi surface disinfectant *
- Antifect® FF *
- B 10 *
Examples of surface disinfectants:
- Cidex® AF
- Gigasept® FF *
* Registered trademarks of the respective manufacturers
The lifespan of the nebulizer chamber with lid and the quartz module is limited to 30
autoclaving cycles. The example disinfectants are recommendations from the device
manufacturer. The recommendation of the mentioned active ingredient bases (pH 4.5-
8) applies. This does not release the user from the need to consult a hygiene specialist
in individual cases or to test the components to be disinfected for compatibility with the
active ingredients.
Sterilization
Certain parts of the device and nebulizer chamber can be steam sterilized.
Steam sterilization 134°C one cycle, 3 minutes duration is allowed only for the
following parts:
Nebulizer Chamber (11)
Nebulizer Chamber lid (36)
Aerosol Tube (15)
Ventilation Tube (16)
Before sterilizing, ensure that any cleaning or disinfection residue is completely
removed. Once sterilized, please examine all parts for possible damage. Any damaged
part should be replaced before use.