6. Operating the device
Nebulization directly from the nebuliser chamber
After the device is assembled use a sterilized nebulizer system and ensure that both,
air filter (33) and bacteria filter (23), are properly installed.
• Medication in doses between 5ml and 10ml can be nebulized.
• Remove the lid (36) from the nebulization chamber (11) by turning it slightly back and
forth and pulling at the same time.
• Plug the conical medication cup (38) on the nebulization chamber lid (36) make sure
the water is between the minimum and maximum levels.
• Place the nebulizing chamber lid to the nebulizing chamber (11). Insert the nebuliser
chamber into the holding device (22) and lock the nebuliser chamber by turning it to
the left until it clicks into place.
• Apply desired medication (max 10ml) into the disposable medication cup (38).
• Connect the aerosol tube (15) and ventilation tube (16) to the chamber lid.
• Turn on the device using the ON/OFF switch which is located on the left side of the
control panel.
• Switch on the device with the mains switch (I). The nebulization process will start.
NOTE!
If the nebulization process does not start immediately, check that the nebulization
chamber is correctly seated in the device holder.
NOTE!
Do not let the nebulizer chamber run dry. Low Level Water alarm: If the water in the
Nebulization Chamber falls below the lower limit, the warning led will illuminate and
the unit will alarm. When water is added, both warning LED and alarm will deactivate.
22