5. MAINTENANCE
5.1 Cleaning
The housing may be cleaned with a cloth dampened with a mild detergent if necessary.
5.2 Troubleshooting
The following table lists common problems and possible causes and remedies.
If the problem persists, contact OHAUS or your authorized dealer.
Symptom
Cannot turn on
Poor accuracy
Cannot calibrate
E E rr 8.1
E E rr 8.2
E E rr 8.3
E E rr 8.4
E E rr 8.5
E E rr 8.6
E E rr 8.7
r r ef.err
Lo.ref
CAL E
USB.Err
Find Quality Products Online at:
WARNING: Electric Shock Hazard. Disconnect the equipment from the power
supply before cleaning.
Attention: Do not use solvents, chemicals, alcohol, ammonia or abrasives to
clean the housing or control panel.
Possible Cause
No power to balance
Improper calibration
Unstable environment
Security switch in locked position
Weight exceeds power on zero range.
Weight below power on zero range.
Over load (weight exceeds rated capacity)
Under load (pan removed)
Tare out of range
Displayed value exceeds 999999 (possible in Totalization mode)
Unknown internal weight position
Parts counting or percentage error – sample weight <1d. Balance
shows error then exits parts counting or goes to [C C lr.APU].
Percent Reference weight or APW is too low for accurate results
Fail to do calibration.
Cannot find menu or app file in U-disk.