Making Measurements
– Your multimeter can display diode forward bias of up to approximately 2.1 V.
NOTE
– If the beeper is enabled during a diode test, the multimeter will emit a beep
Reverse the probes (as shown in Figure 2-14) and measure the voltage across the
diode again. Assess the diode according to the following guidelines:
– A diode is considered good if the multimeter displays OL in reverse bias mode.
– A diode is considered shorted if the multimeter displays approximately 0 V in
– A diode is considered open if the multimeter displays OL in both forward and
Figure 2-12
70
The forward bias of a typical diode is within the range of 0.3 V to 0.8 V;
however, the reading can vary depending on the resistance of other
pathways between the probe tips.
for a normal junction and a continuous beep for a shorted junction below
0.050 V. See "Changing the beep frequency" on page
beeper.
both forward and reverse bias modes, and the multimeter beeps repeatedly.
reverse bias modes.
Open diode display
129