Service & maintenance information
Service and maintenance of a medical device is wholly the responsibility of the owner of that device.
Failure to service and maintain a device in accordance with the instructions may invalidate the
device's warranty. Furthermore, failure to service and maintain a device may compromise the clinical
condition or safety of users and/or their caregivers. Do not perform service and maintenance while the
user is in the product. Contact the local dealer for assistance, if needed, in setting up, using or how to
perform service and maintenance of the device.
Lifetime
The lifetime of this product in normal use is 8
years if all maintenance and servicing is carried
out in accordance with the manufacturer's
instructions and demonstrably recorded. The
device may be reissued within its lifetime.
Service interval
The
frequency
of
inspection
altered depending upon use and wear. It is
recommended that the product is inspected
yearly, every time the product is re-issued for
use and after a long-term storage (more than 4
months). The inspection should be carried out
by a person who understands the use of the
product.
Service checklist:
Check and adjust the following as per their
intended use:
• Seat plane adjustment. The seat plane
is adjustable and stays in position after
adjustment.
• Frame is adjustable in height
• Adjustment options for the foot rest, it is
possible to lock the angle and fasten the
telescopic tubes in position
• Seat back is adjustable and can be locked
in position
• Nuts and bolts on the product are
tightened
• Oil the swing-away parts
• Product for cracks or signs for worn-out
parts
• All labels on the product are intact
Repair or replace any damaged or worn out
parts.
Maintenance
Before every use it is recommended to:
Wipe off any residue and dirt from the product,
using a cloth with warm water and a mild
detergent/soap without chlorine and let it dry
before use carry out a visual inspection, for any
damaged or worn-out parts.
Washing
Main product
This product can be washed at 60° using a
mild detergent for 10 minutes in a washing
machine designed for medical devices. Use the
machine's drying function to dry the product.
Unmount all accessories and wash them
separately.
EN
Before washing
Avoid degreasing of the piston rods, the
actuators should be retracted to minimum
stroke before washing.
• Remove all accessories, unplug all cables
and make sure to dismount the battery,
control-box and handset. Leave the
actuators with the cables mounted on the
may
be
product.
• Do NOT clean the power unit with a high
pressure cleaner or steam cleaner.
After washing
• Ensure correct placement of cables
inserted into the control box when
reinstalling the power unit system.
• Check that the handset functionalities
works correct
Fabric and covers
The cover can be washed at max. 60°C using a
mild detergent in a washing machine.
Separate foam and cover before wash.
Foam
The foam inserts must only be washed by hand.
Use warm water and a mild detergent/soap
without chlorine and let all parts dry before
use.
During and after washing the foam, do not
bend, squeeze or wring it, as it may negatively
affect the shape.
Handwash
This product and its accessories can also be
washed by hand. Use warm water and a mild
detergent/ soap without chlorine and let all
parts dry before use.
Disinfection
The product can be disinfected with a 70%
disinfectant IPA solution. It is recommended to
wipe off any residue and dirt from the product,
using a cloth with warm water and a mild
detergent/soap without chlorine and let it dry
before disinfection.
Materials
• Plastic
• Stainless steel
• Steel
• Polyurethan
• TPE
Surface treatment
The following surface treatment processes