English
Apply A Patch
Please refer to User Guide SECTION 4 for more details.
1.
Wipe the attachment site with an alcohol swab. Wait
until it dries.
2.
Remove the adhesive film from the back of the Patch
and make sure the needle is not sticking out. Tap
Next on the ADM or Narsha.
3.
Apply the Patch. Tap Start Safety Check on the ADM
or Narsha.
4.
Securing the Patch in place, quickly rotate the ne-
edle lever 100° clockwise. Listen for a beep indica-
ting that the cannula is inserted properly.
5.
Turn the lever all the way to completely detach it
from the Patch. The detached area should have a
vertical mark (refer to the image below).
6.
Tap Next on the ADM or Narsha when finished.
Attachment Sites
1. Abdomen
2. Arms & thigh
3. Hip & lower back
Turning Needle Insertion Lever
4
Quick&Troubleshooting Guide PUMP OK.indd 4
Quick&Troubleshooting Guide PUMP OK.indd 4
Begin Insulin Delivery
To create basal or bolus insulin delivery programs plea-
se refer to User Guide SECTION 5 and 6.
•
Once you have activated the Patch, you may now be-
gin the insulin delivery.
•
Create basal/temp basal programs or add bolus/ex-
tended bolus using the ADM or Narsha.
Vertical mark at
detached site
31/12/21 12:12
31/12/21 12:12