Guidance and manufacturer's declaration – electromagnetic
emissions
This device is intended for use in the electromagnetic environment specified below. The user of this device
should make sure it is used in such an environment.
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/Flicker emissions IEC
61000-3-3
Guidance and manufacturer's declaration – electromagnetic
immunity
This device is intended for use in the electromagnetic environment indicated here below. The user of this
device should make sure it is used in such an environment.
IEC 60601
Immunity test
Test level
Electrostatic
±8kV contact
Discharge (ESD)
±15 kV air
IEC 61000-4-2
±2 kV for power supply
Electrical Fast
lines
Transient/Burst
±1 kV for input-output
IEC 61000-4-4
lines
Surge
±1 kV differential mode
IEC 61000-4-5
±2 kV for common mode
DOMU119_EN-FR-ES-RU-DE_v4_20230105_ Horizon® S5
Compliance
The device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
Group 1
likely to cause any interference in nearby electronic
equipment.
Class B
The device is suitable for use in all establishments,
Class A
including domestic establishments and those directly
connected to the public low-voltage power supply
network.
Complies
Compliance level
Electromagnetic environment-guidance
Floors should be wood, concrete or ceramic tile. If floors
±8 kV contact
are covered with synthetic material, the relative humidity
±15kV air
±2 kV for supply mains
Mains power quality should be that of a typical home or
±1 kV for input/output
lines
±1 kV differential mode
Mains power quality should be that of a typical home or
±2 kV for common mode
Electromagnetic environment guidance
should be at least 30%.
hospital environment.
hospital environment.
25