Technical specifications

Specifications

Flow control settings
Rated power (W)
Voltage and frequency
Flow rate (L/min)
Concentration (rated flow)
Sound pressure level dB (A)
Sound power level dB (kg)
Net weight (kg)
Dimensions (mm)
Air outlet pressure
Features
Device operation above or outside of the voltage,
LPM, temperature, humidity and/or altitude
values
specified
may
concentration levels.

When the nominal pressure of the oxygen output
port is zero, the oxygen concentration is 93% ±3%
under the operating environment and rated flow
rate.
See
the
following
concentration and flow rate diagram":
DOMU119_EN-FR-ES-RU-DE_v4_20230105_ Horizon® S5
0,5 to 5 liters
320
220V±22V, 50Hz±1Hz / 110V±11V, 60Hz±1Hz,
depending on the model.
0,5 - 5
93% ± 3%
≤ 45
≤ 55
14,96
380
0.05 ± 10% MPa
Abnormal temperature alarm; low oxygen
content alarm; power failure alarm; timer:
pressure alarm; low flow alarm; thermostat
fault alarm.
decrease
oxygen
"Output
oxygen
240
660
x
x
Relation of outlet oxygen concentration and flow