AC/DC VOLTAGE (LESS THAN 1000V)
1. Insert RED test lead into V jack
COM jack
setting.
NOTE: The meter defaults to AC measurement. To measure DC, press
the "SEL" button
DC icon on the LCD indicates which mode is selected.
2. Apply test leads to the circuit to be tested to measure voltage.
The meter will auto-range to display the measurement in the
most appropriate range.
NOTE: The hazardous voltage indicator
NOTE: If "–" appears on the LCD, the test leads are being applied to
the circuit in reverse. Swap the position of the leads to correct this.
NOTE: When in a voltage setting and the test leads are open,
readings of order mV may appear on the display. This is noise and
is normal. By touching the test leads together to close the circuit
the meter will measure zero volts.
AC/DC LO Z VOLTAGE (LESS THAN 600V)
Voltage measurements in Low Impedance (Lo Z) setting can be used
to identify ghost or stray voltages.
• Follow same procedure for measuring AC/DC Voltage (see above)
with Function Selector switch
DO NOT attempt to measure voltages greater than 600V in
Lo Z setting.