Fujitsu STYLISTIC Q550 Manual Del Usuario página 24

Usage in medical environment
Ocultar thumbs Ver también para STYLISTIC Q550:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 83
6 – English
Table 3: Guidelines and manufacturer's declaration – Electromagnetic
The device or system may be operated in the electromagnetic environment indicated below. The
customer or user of the device or system must ensure that it is used in such an environment.
Immunity
IEC 60601
testing
Test level
Conducted RF
3 V
eff
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
18
immunity
Level of
Electromagnetic environment - guidelines
compliance
Portable and mobile RF communications
equipment should be used no closer to any part of
the device or system, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance:
3 V with
docking unit
and/or
peripheral
cabling, but
without analog
modem
3 V/m
80 to 1000 MHz
3 V/m
1000 to
2500 MHz
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and is the recommended
separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
STYLISTIC Q550
80 MHz to 800 MHz
800 MHz to 2.5 GHz
a
c
Fujitsu Technology Solutions
b
Tabla de contenido
loading

Tabla de contenido