Press shearing head lock 13, see Fig. 4.
12
2
Fig. 5
Fold forward shearing head 12, see Fig. 5.
Detach shearing head 12.
Clean open knife hinge with a brush.
Install shearing head on knife hinge 12.
Press on/off switch 11.
Press shearing head 12 slightly in the direction of
the DURATI battery-powered shearing.
Shearing head 12 will audibly click into position.
Check that the shearing head 12 is correctly seated
on the DURATI battery-powered shearing machine.
5.4.3 Lubricating the shearing head
Clean upper cutting plate and lower cutting plate
with a soft wipe or brush. Make certain the gaps in
serration are cleaned properly.
Slightly lubricate upper cutting plate and lower
cutting plate with special lubricant for shearing
heads GT604.
5.4.4 Li-ion battery
8
3
Fig. 6
Hold DURATI battery-powered shearing machine 2
and press battery release lock 3.
Pull out Li-ion battery 7 from DURATI battery-
powered shearing machine 2.
Insert the charged Li-ion battery 7 in DURATI bat-
tery-powered shearing machine 2 as far as it will
go.
The DURATI battery-powered shearing machine 2
is ready for operation.
6.
Reprocessing
Damage to, or destruction of the
product
cleaning/disinfecting or steriliza-
tion!
CAUTION
Only clean and disinfect the
product manually.
Do not sterilize the product
under any circumstances.
2
3
caused
by
mechanical
9