Guidance And Manufacturer's Declaration - Electromagnetic Emissions - bbluv Ora Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 31
EMC INFORMATION-GUIDANCE AND MANUFACTURE'S
DECLARATION
Do not use near active HF surgical equipment and the RF shielded room of
an ME system for magnetic resonance imaging, where the intensity of EM
disturbances is high.
Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary,
this equipment and the other equipment should be observed to verify that they
are operating normally.
Use of accessories, transducers, and cables other than those specified or
provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 12 in (30 cm) to
any part of the equipment, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could result.

GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC EMISSIONS

EMISSIONS TEST
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations / flicker emissions
IEC 61000-3-3
16
COMPLIANCE
Group 1
Class B
Not applicable
Not applicable
Tabla de contenido
loading

Productos relacionados para bbluv Ora

Este manual también es adecuado para:

B0111Jpd-fr300

Tabla de contenido