TECHNICAL SPECIFICATIONS
Product Name
Product Model
Power Supply Mode
Operating Voltage
Battery Model
Operating Mode
Display
Measure Time
Latency Time
Measuring Range
Accuracy
(Laboratory)
Accuracy (Clinical)
Resolution
Memory
Low-Battery Alert
Automatic Power-off
Outer dimensions
Weight
Operating
Environment
THERMOMETER
Orä
B0111 / JPD-FR300
Internal Power Supply
DC 3V
AAA x 2
Continuous operating
Segment LCD
About 1 second
About 3 seconds
Forehead Mode: 22.0 °C - 43.0 °C (71.6 °F - 109.4 °F)
Ear Mode: 34.0 °C - 43.0 °C (93.2 °F - 109.4 °F)
Object Mode: 0 °C - 100 °C (32 °F - 212.0 °F)
Forehead Mode: ± 0.2 °C (36 °C - 39 °C)
± 0.3 °C (22 °C - 36 °C / 39 °C - 43 °C)
Ear Mode: ± 0.2 °C (36 °C - 39 °C)
± 0.3 °C (34 °C - 36 °C / 39 °C - 43 °C)
Object Mode: ± 1 °C (± 2 °F)
± 0.3 °C (± 0.6 °F)
0.1 °C (0.1 °F)
20 temperature readings
The low-battery symbol is displayed if the power
voltage is lower than 2.4 V±0.1V
The thermometer automatically powers off if it is
not used in 10±1 seconds.
155.9 x 40.2 x 49.2 mm
90 g (with batteries)
Temperature: 10°C~ 40°C (50°F–104°F)
Humidity: 15%–95% RH, non-condensing
Atmospheric pressure: 86–106 kPa
Orä