Explanation Of Functions - Omron HEM-907 Manual De Instrucciones

Ocultar thumbs Ver también para HEM-907:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 29
Names and Functions of the Parts

Explanation of Functions

(1) Pressure setting function
Inflation can be set by AUTO (automatic setting) or the target value.
AUTO (automatic setting): In the SINGLE (single measurement), AVG. (average), and MANU. (auscultation) Modes, the
monitor estimates the systolic blood pressure value during inflation and automatically inflates to the proper value.
Target value setting: The monitor inflates to the target value.Pressure value is set to 30 to 40 mmHg above the expected
systolic pressure.
(2) Non-display function
A function not to display the results of measured blood pressures.
However, the pressure values during the measurement are displayed.
This function can be used when SINGLE (single measurement) or AVG.
(average) Mode is selected.
(3) Manual deflation control function
A function to accelerate the deflation speed by pushing a DEFLATION Button
during deflation.
This function can be used when MANU. (auscultation) Mode is selected.
6
P - SET
HIDE
ON/OFF
START
HIDE
P-SET
MODE
SINGLE
140
180
AVG.
MANU.
220
CHECK
100
STOP
260
280
AUTO
ON/OFF
START
HIDE
P-SET
MODE
SINGLE
140
180
AVG.
MANU.
220
CHECK
100
STOP
260
AUTO
280
Tabla de contenido

Solución de problemas

loading

Tabla de contenido