Rs232 (Db9) Pin Connections - OHAUS Pioneer PA124 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

EN-11
6.3 Communication
6.3.1 Commands
The RS232 Interface allows a computer to control the balance as well as receiving data such as the displayed weight.
The balance will return "ES" for invalid commands.
Command Function
IP
Immediate Print of displayed weight.
P
Print displayed weight (uses Stable ON/OFF menu settings).
CP
Continuous Print.
xP
Interval Print x = Print Interval (1-3600 sec)
T
Same as pressing Zero Key.
ON
Turns balance ON.
OFF
Turns balance OFF.
PSN
Show Serial Number.
PV
Version: Print product name, software revision and LFT ON (if LFT is set ON).
PU
Print current mode/unit
x#
Set PC ref wt (x) in grams
P#
Print PC ref wt
x%
Set % ref wt (x) in grams
P%
Print % ref wt

6.3.2 RS232 (DB9) Pin Connections

Pin 2: Balance transmit line (TxD)
Pin 3: Balance receive line (RxD)
Pin 5: Ground signal (GND)
Pin 7: Clear to send (hardware handshake) (CTS)
Pin 8: Request to send (hardware handshake) (RTS)
6.4 Compliance
Compliance to the following standards is indicated by the corresponding mark on the product.
Mark
Standard
This product complies with the applicable harmonized standards of EU Directives 2011/65/EU
(RoHS), 2014/30/EU (EMC), 2014/35/EU (LVD) and 2014/31/EU (NAWI). The EU Declaration of
Conformity is available online at www.ohaus.com/ce.
EN 61326-1
CAN/CSA-C22.2 No. 1010.1
UL Std. No. 3101-1
RxD
GND
TxD
5
9
6
RTS
Pioneer
1
CTS
Tabla de contenido