Cleaning And Disinfecting The Doppler; Cleaning And Disinfecting Probes; Specifi Cations; Equipment Classifi Cation - Huntleigh FD1 Instrucciones De Uso

Ocultar thumbs Ver también para FD1:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 38
EN
7.2

Cleaning and Disinfecting the Doppler

Always keep the external surfaces clean and free of dirt and fl uids using a clean dry cloth.
Wipe any fl uids from the surface of the product using a clean dry cloth.
Wipe with a cloth dampened in 70% Isopropyl Alcohol.
Completely dry with a clean, dry cloth.
If the product has been contaminated use the methods described for probes
7.3

Cleaning and Disinfecting Probes

Clean the probes before examining a patient using the low risk cleaning method below. Following patient
examination, clean and/or disinfect the probes by the appropriate method based upon the level of cross
contamination risk, as defi ned below:
Risk
Defi nitions
Low
Normal use or low risk situations
include patients having intact
skin and no known infection.
Medium
The patient has a known
infection, skin is not intact, the
part is heavily soiled.
High
This procedure should only be
used when the part has been
contaminated by blood.
Caution
Repeated and unnecessary use of concentrated solutions will result in damage to the product. Do not
allow Sodium Hypochlorite solutions to come into contact with metal parts.
The use of disinfectant materials other than those listed is the responsibility of the user for their effi cacy and
compatibility with the device.
8.
Specifi cations
8.1
Equipment Classifi cation
Type of protection against electric shock
Degree of protection against electric shock
Mode of operation.
Degree of protection against harmful ingress of
particles and/or water.
Degree of safety of application in the presence of
a fl ammable anaesthetic
*For home use, this can be upgraded to IPX2 when using the protective pouch (ACC-OBS-080).
Procedure
1. Remove soiling, wipe with a mild neutral detergent and
then wipe with a cloth dampened in water.
2. Completely dry with a clean cloth.
1. Follow low risk procedure then wipe with a cloth
dampened in Sodium Hypochlorite (1,000ppm).
2. After two minutes wipe with a cloth dampened in water
and then dry with a clean cloth.
1. Follow low risk procedure then wipe with a cloth
dampened in Sodium Hypochlorite (10,000ppm).
2. After two minutes wipe with a cloth dampened in water
and then dry with a clean cloth.
Internally powered equipment
Type BF - equipment with an applied part
Continuous
Main Unit: IP20*,
Probes : IPX7
Equipment not suitable for use in the presence of a
FLAMMABLE ANAESTHETIC MIXTURE WITH AIR,
OXYGEN OR NITROUS OXIDE
Tabla de contenido
loading

Este manual también es adecuado para:

Fd3

Tabla de contenido