CLEANING & DISINFECTION INSTRUCTIONS FOR SINGLE PATIENT USE
WARNING
To prevent possible risk of infection from contaminated cleaning/disinfection solutions, always prepare fresh solution for each cleaning cycle and discard solution after each use.
Containers
IMPORTANT
1.
The disposable collection container for single-patient use only.
2.
The reusable collection container can be autoclaved between patients and reused.
All Containers
1.
Turn unit Off and allow vacuum to drop.
2.
Disconnect from power source.
3.
Disconnect tubing and remove container from holder.
4.
Carefully remove lid and empty contents.
5.
Disassemble and wash/disinfect container following appropriate instructions below.
NOTE– The disassembled disposable collection container may also be washed in a
dishwasher, top shelf only, using a cycle with a water temperature between 55˚C to 65˚C
(131˚F to 149˚F).
Disposable w/Internal Filter Cartridge
1.
Remove filter cartridge and short connection tubing and set aside.
2.
Wash container, lid, and splash guard in warm water/dishwashing solution. Rinse with
clean warm water, then disinfect.
3.
After washing, disinfect in 1 part vinegar (>+5% acetic acid concentration) to 3 parts
water (55˚C to 65˚C / 131˚F to 149˚F) solution for 60 minutes. Rinse with clean warm
water and air dry.
WARNING
Do not remove float ball from lid. If removed, float ball may
pose a choking hazard.
IMPORTANT
Filter MUST NOT get wet. The filter material cannot be removed from the cartridge.
Disposable w/External Bacteria Filter
1.
Remove bacteria filter, elbow, and short connection tubing and set aside.
Wash container and lid in warm water/dishwashing solution. Rinse with clean warm
2.
water, then disinfect.
3.
After washing, disinfect in 1 part vinegar (>+5% acetic acid concentration) to 3 parts
water (55˚C to 65˚C / 131˚F to 149˚F) solution for 60 minutes. Rinse with clean warm
water and air dry.
Reusable Container with External Bacteria Filter
1.
Remove bacteria filter, elbow, and short connection tubing and set aside.
2.
Wash container, lid, o-ring, and overflow valve in a solution of warm water with a mild,
liquid detergent (e.g. Dawn or Palmolive) and rinse with clean, warm tap water.
3.
Then disinfect using one of the following methods:
Disinfection for single patient use:
a. All Containers: After washing, disinfect in 1 part vinegar (>+5% acetic acid
concentration) to 3 parts water (55˚C to 65˚C / 131˚F to 149˚F) solution for 60
minutes. Rinse with clean, warm water and air dry in a clean environment.
b. Only Reusable Container: Soak with a commercial (bacterial-germicidal)
disinfectant. Follow disinfectant manufacturer's recommended dilution rates and
instructions carefully.
A-7325
EN - 13