Error Messages - KERN FFN-N Serie Manual De Instrucciones

Balanza compacta
Ocultar thumbs Ver también para FFN-N Serie:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 64

10 Error messages

Display
29
Description
Zero range exceeded
A/D converter outside range
Displayed weight is
permanently changing
Overload
Remedy
Unload the balance
Unload the balance; check if
the weighing is correctly
placed and screwed on
Avoid air draughts / air
movement as well as
vibrations of the table and of
the floor
Unload the balance and
adjust again
FFN-N-BA-e-1121
Tabla de contenido
loading

Tabla de contenido