Control Panel - Elmi CM-8S Manual De Usuario

Centrifugadora de mesa
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Open the lid by pressing the OPEN button .
Check the rotor, make sure that rotor nut is se-
curely fastened . Try rotating it by hand . There
should be no noise or obstacles for smooth run-
ning .
Check the adapters . All of the adapters must be
ATTENTION! If there is any apparent damages or malfunctions, DO NOT TURN ON the centrifuge without consult-
ing a specialist.
ATTENTION! To finish the unpacking of centrifuge, pull the emergency opening lever to the side (as shown on the
figure on page 7) and remove all packaging materials from the rotor bucket.
Centrifuge consists of a stainless still casing, rotor
mounted in it, electric motor and control system .
The rotor is covered with self locking transparent
lid . A network switch is mounted in the back of the
Adjustable parameter
Parameter value
Tuning keys
Settings menu key

Control panel

Centrifuge CM-8S
equally positioned inside the brackets .
Check the device's immediate surroundings .
A safety distance of 30cm must be observed
around the centrifuge during operation .
Check for other signs of damage or malfunc-
tion .
casing . A control panel on the front of the casing
contains display and buttons with the following
functions:
Scroll keys
Lid lock indicator (green - opened, red - closed)
Braking level indicator
Display brightness, beeper volume and braking
level can be adjusted in the Settings menu
USER MANUAL
Button to unlock the lid of centri-
fuge .
Button to start the centrifugation
or to end the cycle before the es-
timated time .
5
ENG