5. ADVANCED OPTIONS

The handheld allows the user to view, edit and set some operating parameters by simply combining the keys
provided. Follow the steps given below to access:
KEY COMBINATION
+
+
+

5.1. SETTING THE SAFETY UNLOCK MODE

The myray RX DC - RX DC/I x-ray unit has a safety unlock for the ball joint.
The default setting allows the ball joint to be disengaged by simply touching one of the keys present on the front of
the head. To prevent accidental contact with the keys from unexpectedly disengaging the ball joint (and therefore
causing undesired movement of the head), the safety unlock mode can be activated. In this mode, the ball joint is
disengaged only if both keys are activated at the same time.
To set the safety unlock mode, press keys
Scroll the parameters up to parameter P05 and press key
.
Press key
to quit the set up menu.
72
Press these two keys to adjust the sensitivity levels (determined based on the table
given below and type of sensor/receiver used), modifying the current value from
the minimum to the maximum allowable one (on a scale from 3 to 25), with keys "+"
and "-".Press key "size" to confirm the desired level and go back to the main
screen.
This menu is not available in USER mode.
Hold down these two keys to go to the set up menu (from P 01 to P 07).
Press key "Build" to make the selection. Once within the individual configurations,
they can be scrolled with keys "+" and "-" and selected by pressing key "Build"
again. Key "tooth" quits set up without saving the setting.
The configurations are given in detail below:
-
P 01: Sets the stand by time (from a minimum of 5 to a maximum of 30
minutes).
-
P 02: Assigns an identification tag to the x-ray unit's base (from 1 to 5 or none).
-
P 03: Shows the list of software versions.
-
P 04: Handheld code display.
P 05: Activates/deactivates the safety unlock mode (see section 5.1). (only RX
DC Plus).
-
P 05: Reserved. (only RX DC eXTend).
-
P 06: Selects the operating mode (En60, En63, En65 and AUTO).
-
P 07: Sets the type of removable cone used.
Activating/deactivating the USER mode. Icon
is activated.
and
OPERATOR'S MANUAL
DESCRIPTION OF COMMAND
to go to the set up menu.
. Scroll the options to select "ON" and press key
comes on to signal USER mode
EN