Gima OXY-100 Manual De Uso Y Mantenimiento página 28

Oxímetro
Ocultar thumbs Ver también para OXY-100:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

ENGLISH
28
4.4.1 Setting
On the menu screen, select "Setting" and then press "
" key for entering
into system setup screen. The setup screen is as shown in the following figures.
Figure 4-4 System Setup Screen
Operation Instructions:
- Patient: according to the type of SpO
probe, the matched patient type should
2
be selected, there are two options:"ADU" for adult, and "NEO" for neonate.
- SpO
Lo-Limit: SpO
low limit setting; range: 50%~99%, the step is 1%. The
2
2
factory default value for adult is 90% and 85% for neonate.
- PR Hi-Limit: High limit setting of pulse rate; range: 100~240bpm. From 100
to 150, the step is 1bpm, and from 150 to 240, the step is 5bpm. The factory
default value for adult is 120bpm and 160bpm for neonate.
- PR Lo-Limit: Low limit setting of pulse rate; range: 30~99bpm, and the step
is 1bpm. The factory default value for adult is 50bpm and 60bpm for neonate.
- Date: Date setting
When cursor stays on the Year of the date, press "
" (Mode/OK) key to
active Year option, the cursor flashes on the Year of the date;
Press "
/
" (Navigation key) to adjust year.
Press "
" (Power/back) key or "
" (Mode/OK) key to confirm and exit
from date setting.
The procedures of adjusting Month value and Day value are the same with
Year adjustment.
Date Format: yy-mm-dd
Note: The setting operations of other parameters (such as TIME, PATIENT,
RECORDING INTERVAL, POWER SAVING etc.) are the same with date
setting.
Tabla de contenido

Solución de problemas

loading

Tabla de contenido