Operating advice

1
> Fill flask n°2 with clean water
(tap water).
Figure
1.
4
> Immerse the sensor
in flask n°1 (pH 7).
Figure
4.
28
All manuals and user guides at all-guides.com
3.1.

Calibration procedure

Refer to paragraph 4.2. to change the language.
!
Important:
The sensor must be calibrated when your pH Perfect is first installed and
ideally every two months there after.
> Calibration must be done each time your pool is restarted at the beginning of
the season.
> The procedure should be repeated if an abnormal amount of corrective solution
is consumed (pH minus or pH plus).
> The calibration procedure can only be carried out when filtration is stopped.
> Before calibrating, stop filtration and close the valves so as to isolate the pH
sensor holder from flowing water.
2
> Undo the sensor holder and
remove the sensor from the pipe.
Figure
2.
5
a
b
3
> Rinse the sensor in flask n°2 (water).
Then shake gently to remove
excess water.
Figure
3.
c
> Keep the "CAL/Enter" button pressed
for more than 5 seconds until
"Calibration" appears on the
LCD screen
(Figure
5a).
Press the "CAL/Enter" button again
as indicated on the screen for
calibration on pH7
(Figure
5b):
an automatic progression bar then
appears
(Figure
5c). Keep the
sensor immersed in the buffer
solution for 30 seconds.