7
O2 Calibration
(PaTH: H \ Cal \ Calibrate Sensor)
Select method 1-point
h
14
Choose "SLOPE" or "OFFSET" calibration through pressing the corresponding field.
Press the "CaL" button for starting calibration.
Place the sensor in air or the calibration gas and press "nEXT" button
Enter the value for the calibration point (Point1).
The M800 checks the stability of the measuring signal and proceeds as soon as the
signal is sufficiently stable.
NOTE: If option "Stability" is set to "Manual" press "nEXT" after the measuring signal
is stable enough to go on with the calibration.
The transmitter shows the value for the slope and the offset* as the result of the
calibration.
The calibration values are stored in the calibration history and taken over (button
"aDJUST"), stored in the cal history and not taken over (button "CaLIBRaTE") or
discarded (button "CanCEL").
* For amperometric sensors P100 and P0 for optical sensors.