Select method Process
pH Process calibration
Press the "CaL" button to start calibration.
Take a sample and press the e button to store the current measuring value. To show
the ongoing calibration process, "P" is blinking in the Start and Menu Screen if the
related channel is selected in the display.
after determining the pH value of the sample, press the c icon in the Menu Screen
again.
Enter the pH value of the sample. Press the "nEXT" button to start the calculation of
the calibration results.
The display shows the value for the slope and the offset as the result of the calibration.
The calibration values are stored in the calibration history and taken over (button
"aDJUST"), stored in the cal history and not taken over (button "CaLIBRaTE") or
discarded (press button "CanCEL").