Safety Instructions - Dr.Mach Mach LED 130 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 14
EN
User manual - English
EC conformity symbol
This symbol indicates that this is a medical
device
Unique device identifier (ID) of the product
NRTL Test mark
The lamp was tested by a 'Nationally Rec-
ognized Testing Laboratory'
Reference to China RoHS /
pollution control logo
Instruction for disposal of the device

Safety instructions

The instructions for use must be observed for safe
handling of the lamp
To avoid the risk of an electric shock, this device
must only be connected to a supply network that
has a protective earthing conductor.
A primary-side ON/OFF switch to isolate the sys-
tem from the supply network must be provided on
site. The switch must meet the requirements of
IEC 61058-1 for nominal voltage peaks of 4 kV.
This device is not designed for operation in envi-
ronments enriched with oxygen.
59141002
The lamp may only be used for the intended pur-
pose. Otherwise, the manufacturer will not be lia-
ble for personal injury or damage to property.
The lamp is equipped with a sterilisable handle at
the factory and must only be used with this handle.
Changes to the light are prohibited and will invali-
date the manufacturer's certificate of conformity
and all warranty claims.
Use only the mains units (or transformers) ap-
proved or supplied by the manufacturer. Non-ob-
servance will void the conformity of the product
and release the manufacturer from any claims un-
der warranty.
Installation, maintenance and repair work may only
be carried out by the manufacturer or by specially
trained staff.
Maintenance must be carried out on the light at
least every two years
Additional equipment that is connected to medical
electrical equipment, must conform to the relevant
IEC- or ISO standards (e.g. IEC 60950 or IEC
62368 for data processing equipment). Moreover,
all configurations must meet the requirements for
medical electrical systems (see Section 16 of the
latest version of IEC 60601-1). Anyone who con-
nects additional equipment to medical electrical
equipment, is configuring a medical system and is
therefore responsible for ensuring that the system
meets the requirements for medical electrical sys-
tems. In case of doubt, contact your local repre-
sentative or our technical customer services.
6 / 70
The simultaneous use of several lights to illumi-
nate a wound area may result in the maximum al-
lowed energy input being exceeded (1,000 W/m
and thus excessive heat development. It is the
user's responsibility to ensure that the maximum
allowed limit is not exceeded.
The unprotected human eye can be damaged by
direct light. Do not look directly into the light beam
of the lamp. Do not point the light beam at the pa-
tient's unprotected eye continuously.
When positioning the light body, there is a risk of
injury (e.g. crushing) and collisions with other ob-
jects (inventory) or walls.
Parts that fall off could injure the patient or lead to
an infection of the wound area.
Do not remove the rating plate or the warning la-
bels.
It is forbidden to carry out servicing or repair activi-
ties whilst the lamp is in use.
Mach LED 130
2
)
V1
2021-05-07