Etac R82 Wombat Living 1 Guia Del Usuario página 40

Tabla de contenido
Test modes during immunity tests:
Battery mode, standby armed/ready.
Charging mode, actuator not running.
EM immunity standards
Electrostatic discharge
EN
IEC61000-4-2
Radiated RF EM fields
IEC 61000-4-3
Proximity fields form RF wireless
communication equipment
IEC 61000-4-3
Electrical fast transients / bursts
IEC 61000-4-4
Surges, Line-to-line
IEC 61000-4-5
Conducted disturbance induced
by RF fields
IEC 61000-4-6
Rated
power
magnetic fields
IEC 61000-4-8
Voltage dips, short interruptions
and voltage variations
IEC 61000-4-11
Variation of power frequency
IEC 61000-4-28
IEC 60601-1-2:2014 test levels
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
IEC 60601-1-2:2014 Table 9
± 1 kV, ±2 kV
100 kHz repetition frequency
± 0,5 kV, ± 1 kV
230 VAC
6 Vrms
0.15 MHz – 80 MHz
80% AM at 1 kHz
frequency
30 A/m
50 Hz
0 % U
for 0,5 cycle at 0°, 45°,
T
90°, 135°, 180°, 225°, 270°
and 315° phase angles
0 % U
for 1 cycle
T
70 % U
for 25 cycles
T
0 % U
for 250 cycles
T
230 VAC
N/A
Additional test levels
± 6 kV contact
20 V/m
800 MHz – 2.5 GHz
80 % AM at 1 kHz
N/A
± 1 kV, ± 2 kV
5 kHz repetition frequency
± 0,5 kV, ± 1 kV
100 VAC and 240 VAC
N/A
30 A/m
50 Hz and 60 Hz
0 % U
for 1 cycle
T
40 % U
for 10 cycles
T
70 % U
for 25 cycles
T
100 VAC and 240 VAC
49 Hz and 51 Hz
100 VAC and 240 VAC, 50 Hz
59 Hz and 61 Hz
100 VAC and 240 VAC, 60 Hz
support.R82.org
40
Tabla de contenido
loading

Este manual también es adecuado para:

R82 wombat living 2R82 wombat living 3

Tabla de contenido