Guidance and manufacture's declaration – electromagnetic immunity – for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture's declaration – electromagnetic immunity
The JC-119M (TMB-1112) is intended for use in the electromagnetic environment specified below.
The customer or the user should assure that it is used in such an environment.
IEC 60601
Immunity test
test level
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 V/m 80 MHz
Radiated RF
to 2.5 GHz
IEC 61000-4-3
Compliance
Electromagnetic environment -
level
guidance
Portable and mobile RF communications equipment should be
used no closer to any part of the YS-6100, including cables, than
the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Portable and mobile RF communications equipment should be
used no closer to any part of the including cables, than the
recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.167
d = 1.167
3 Vrms
d = 2.333
where P is the maximum output power rating of the transmitter
3 V/m
in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey, a should be less than the
compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with
the following symbol:
64
80 MHz to 800 MHz
800 MHz to 2.5 GHz