4.
Confirm System Check Screen .
•
This display appears when you turn on the meter . Do not use the meter if
the display check screen does not exactly match the example. Contact
Customer Care.
•
See The FreeStyle Freedom Lite® Meter and How It Works section for more
information .
•
If the system check screen does not appear, go to Troubleshooting section .
5.
Apply and hold control solution to 1 sample area of the test strip until:
•
You see short lines moving clockwise on the display or
•
You hear a beep
This indicates the test strip obtained enough control solution and the meter is
processing . If the test does not start after applying the control solution, go to
Troubleshooting section .
IMPORTANT: If a short line does not appear after 5 seconds, the
sample may be too small . Do not add control solution to both sides
of the strip . The test strips are for single use only . Do not reuse test
strips . Discard used test strips .
26