5 SPECIFICATION
Enclosure material
Dimension
Weight
Ingress protection
Cable entry
Power
Electrical output
Relays:
Sounder Out:
Operating temperature
Humidity
Repeatability
Zero drift
Approval codes
Standards:
Zones
EMC compliance
26
ADC 12 aluminium alloy
156 x 166 x 109mm
(6.1 x 6.5 x 4.3inch)
Aluminium alloy 1kg (2.2lbs)
IP65 & IP66 (with weatherproof cap)
2x M20 (stopping plug fitted to left-side entry) or supplied with
½" NPT adapters.
10-30Vdc. 3W max
4-20mA current sink or source
RS-485 Modbus RTU
HART (optional)
Alarm 1, Alarm 2, Fault
SPDT contacts rated 1A 24Vdc
MOSFET open-collector drive.
24Vdc (nominally), 250mA maximum load
-40°C to +70°C (-40°F to 158°F)
Note: sensor operating temperatures vary.
Refer to the sensor module datasheet or contact
Crowcon for specific sensor data.
0 to 95% RH, non-condensing
+/- 2% FSD
+/- 2% FSD per year maximum
ATEX and IECEx
Ex II 2G Ex db IIC T6 Gb
Ex II 2D Ex tb IIIC T80°C Db
Certificate numbers:
TUV 16 ATEX 7908 X
IECEx TUR 16.0035 X
EN60079-0:2012 + A11:2013
EN60079-1:2014
EN60079-31:2014
IEC60079-0:2017 Edition 7
IEC60079-1:2014-06
IEC60079-31:2013
Certified for use in Zone 1 and Zone 2 areas
EN50270:2015
Xgard Bright