Recommended Naída CI Q70 Operating and Storage Temperatures
Operating and Storage Temperature Ranges
Condition
Operating temperature
Storage temperature
Approved Power Sources
Advanced Bionics has verified the Naída CI Q70 to function with the power sources listed
below; behavior when using power sources other than those listed here cannot be guaranteed.
Approved Naída CI Q70 Power Sources
Power Source
PowerCel
PowerCel Mini
Zn-Air
Programming Interface
(used during fitting, only)
Approved Naída CI Q70 Power Supplies
Power Supply
PowerCel
Zn-Air
Programming Interface
Troubleshooting the Naída CI Q70
If no sound is heard or the red LED is flashing once per second on the Naída CI Q70 processor
LED when the headpiece is in place over the implant, try the following troubleshooting steps:
Visually inspect the cable for any damage or breakage and verify it is firmly attached to
1
the headpiece and sound processor.
Remove any materials (hat, scarf, headband, etc.) that may be covering the microphone.
2
Unaided listeners may perform a listening check of the microphones as described in the
3
Listening to the Microphone and Auxiliary Input
Minimum
0°C (32°F)
-20°C (-4°F)
Model Number
CI-5517-XXX
CI-5523-XXX
CI-5521-XXX
CI-5527-XXX
PowerOne p675
CI-6101
Rated Voltage
3.7 V
2.8 V
5.0 V
Type
Li-ion (Rechargeable)
Zn-Air (Disposable)
DC
section of the User Guide.
Naída CI Q70 Sound Processor User Guide
Maximum
45°C (115°F)
55°C (131°F)
Capacity (mAhr)
170
230
110
170
1,140
Not Applicable
Rated Energy
0.4 Wh – 0.9 Wh
1,240 mWh
1,250 mW