4 Insert the plug of the AC adapter in a power outlet that is easily accessible.
ð To remove the connector plug, push the release button and withdraw the connector plug.
4.5 Switching the instrument on and off
Switching the instrument on
1 Press the On/Off button.
ð The power button lights up green.
2 The instrument is initialized and a system test is performed. This process will take some time.
ð If only one user is defined, the instrument starts directly with the home screen without showing the login
screen.
Switching the instrument off
1 Tap Home > Log out > Shut down.
- or -
Press the On/Off button.
2 The instrument stops running tasks and shuts down. This process will take some time.
ð When the light goes out, the instrument is in the standby mode.
Note
In the standby mode, the control circuit for the On/Off switch is energized. The rest of the instrument is no
•
longer energized.
4.6 Connecting sensors
Connect the sensor and make sure that the plugs are properly inserted. For sensors with a built-in temperature
probe, connect the second cable to the appropriate ATC socket (ATC int). When using seperate temperature
sensors, connect the cable to the appropriate ATC socket (ATC ext).Twist the RCA (Cinch) plug to ease the
attachment of the sensor.
®
ISM
sensor
®
ISM
sensors are automatically recognized upon connection.
When connecting an ISM
running, the sensor identification will be displayed after the endpoint of the analysis is reached.
4.7 Connecting autosampler
InMotion autosampler
With the InMotion autosampler the automatic measurement of up to 303 samples is possible. InMotion is only
supported when the instrument is connected to the LabX software. The InMotion autosampler is connected to the
instrument with an USB cable and can only be used in LabX connected mode. Several different configurations are
available for the InMotion autosampler depending on your requirements.
Rondolino sample changer
With the Rondolino sample changer the automatic measurement of up to 9 samples is possible. To connect the
Rondolio sample changer to the instument, a "Rondolino USB-TTL box" is required. Use the USB cable to connect
the USB A interface of the instrument to the USB-TTL box. Use the 9-pin cable to connect the the Rondolino
sample changer with the TTL box. Connect the compact stirrer to the stirrer adapter cable and then to the
instrument. The compact stirrer is an overhead stirrer.
4.8 Connecting stirrers
Connect uMix™, the METTLER TOLEDO external magnetic stirrer, or the METTLER TOLEDO Compact Stirrer
(overhead stirrer, adapter cable needed) to the instrument. The stirrers are powered by the instrument and will be
automatically switched on/off during an analysis according to the settings. Before connecting a stirrer, switch off
the instrument. Use the stirrer socket (Stir) on the rear side, note the designation. Note the corresponding chapter
of the operating instructions concerning the settings of peripheral devices.
20
Putting into Operation
®
sensor to the instrument, verify that no analysis is running. In case an analysis is
SevenExcellence™