CLEANING, SANITIZATION, DISINFECTION AND STERILIZATION
Turn off the unit before any cleaning and disconnect the power cord from the mains.
CLEANING THE DEVICE AND THE TUBE'S OUTER SURFACE
Only use a cloth dampened with antibacterial detergent (non-abrasive and free of solvents
of any kind).
SANITIZATION AND DISINFECTION OF THE ACCESSORIES
We recommend a personal use of both the nebulizer and accessories to avoid any risk of
infection or contagion. Open the nebulizer by turning the top (C6.1) counterclockwise,
then remove the nozzle from the top (C6.1), and disassemble it as shown in the "Connection
Diagram" section C6.
SANITIZATION
Before and after each use, sanitize the nebulizer and accessories according to one of the
methods described below.
(Method A): The accessories that can be sanitized with the method A are (C1-C2-
C3-C4-C5-C6.1-C6.2-C6.3-C6.4). Sanitize the accessories in warm (approximately 40° C)
drinking water with mild dishwashing liquid (non abrasive) or in the dishwasher with hot
cycle.
(Method B): The accessories that can be sanitized with the method B are (C1-C2-
C3-C4-C5-C6.1-C6.2-C6.3-C6.4). Sanitize the accessories by immersion in a solution
consisting of 50% water and 50% white vinegar.
Finally rinse thoroughly with warm (approximately 40° C) drinking water.
(Method C): The accessories that can be sanitized with the method C are (C3-C4-C5-
C6.1-C6.2-C6.3-C6.4). Sanitize the accessories by boiling in water for 20 minutes; it is
preferable the use of demineralized or distilled water to avoid scale deposits.
After sanitization shake the accessories vigorously and place them on a paper towel, or
alternatively, dry them with a jet of hot air (e.g. using a hair dryer).
DISINFECTION
The accessories that can be disinfected are (C1-C2-C3-C4-C5-C6.1-C6.2-C6.3-C6.4).
The disinfection procedure described in this paragraph is to be performed before using
the accessories and is effective only if all the prescriptions herein are met and all the
components involved are previously sanitized.
The disinfectant to be used must be of electrolytic oxidizing chlorine type, (active
ingredient: sodium hypochlorite), specific for disinfection and is available in all pharmacies.
Procedure:
- Fill a container, suitable to contain all the individual components to be disinfected, with
a solution of drinking water and disinfectant, while respecting the proportions indicated
on the packaging of the disinfectant itself.
- Completely immerse each individual component in the solution, taking care to avoid
the formation of air bubbles in contact with the components. Leave the components
immersed for the period of time indicated on the packaging of the disinfectant, and
associated with the concentration chosen to prepare the solution.
- Remove the components now disinfected and rinse thoroughly with lukewarm drinking
water.
- After disinfection shake the accessories vigorously and place them on a paper towel, or
alternatively, dry them with a jet of hot air (e.g. using a hair dryer).
- Dispose of the solution as recommended by the disinfectant manufacturer.
STERILIZATION
The accessories that can be sterilized are (C3-C4-C5-C6.1-C6.2-C6.3-C6.4).
The sterilization procedure described in this paragraph is effective only if all the