Table 10.2
Guidelines and Manufacturer Declaration – Electromagnetic Interference Immunity
Guidelines and Manufacturer Declaration – Electromagnetic Interference Immunity
The KaWe examination lamp is designed to be used in the types of environments listed below.
The customer or user of the KaWe examination lamp is responsible for ensuring that this device is used in such an environment.
Interference
immunity test
Electrostatic
discharge (ESD)
according to
IEC 61000-4-2
Fast transient electrical
disturbances/ bursts according
to IEC 61000-4-4
Surges according to
IEC 61000-4-5
Voltage dips, short inter-
ruptions
and voltage variations
on power supply input lines
according to IEC 61000-4-11
Magnetic field with a power
frequency of (50/60 Hz)
according to IEC 61000-4-8
Note: U
is the mains voltage prior to application of the test level.
T
All manuals and user guides at all-guides.com
IEC 60601-
test level
± 6kV contact discharge
± 8 kV air discharge
± 2kV for power supply lines
± 1kV for input and output
lines
± 1 kV differential mode
voltage
± 2 kV common mode
voltage
< 5 % U
T
(>95 % dip in the U
)
T
for a ½ period
40 % U
T
(60 % dip in the U
)
T
for 5 periods
70 % U
T
(30 % dip in the U
)
T
for 25 periods
< 5 % U
T
(>95 % dip in the U
)
T
for 5 seconds
3 A/m
Compliance
level
± 6kV contact discharge
± 8 kV air discharge
± 2kV for power supply lines
Not applicable
± 1 kV differential mode
voltage
± 2 kV common mode
voltage
< 5 % U
T
(>95 % dip in the U
)
T
for a ½ period
40 % U
T
(60 % dip in the U
)
T
for 5 periods
70 % U
T
(30 % dip in the U
)
T
for 25 periods
< 5 % U
T
(>95 % dip in the U
)
T
for 5 seconds
30 A/m
Electromagnetic environment –
guidelines
The flooring should be wood or
concrete or be covered with ceramic tile.
If the floor material is a non-conductive,
synthetic material, the relative humidity
of the air must be at least 30%.
The quality of the mains supply voltage
should be the same as that of a typical
commercial or hospital environment.
The quality of the mains supply voltage
should be the same as that of a typical
commercial or hospital environment.
The quality of the mains supply voltage
should be the same as that of a typical
commercial or hospital environment.
If the user of the KaWe examination lamp
requires continued operation during
power interruptions, it is recommended
that the KaWe examination lamp be
powered from an uninterruptable power
supply or a battery.
Magnetic fields with their line
frequencies should be the same as those
typical of commercial and hospital
environments.
19