6.2 Adjusting the arm
• adjust the clamping lever
If the light body is hard to move or does not stay in the desired posi-
tion, the braking action of the clamping lever must be adjusted.
7. TECHNICAL DATA
7.1 Photometric data
Central illumination intensity
at a distance of 0.5 meters
Central illumination intensity
at a distance of 1 meter
Light spot diameter d
at a distance of 0.5 m
10
Light spot diameter d
at a distance of 1 m
10
Light spot diameter d
at a distance of 0.5 m
50
Light spot diameter d
at a distance of 1 m
50
Residual illumination intensity with one shutter
Residual illumination intensity with two shutters
Residual illumination intensity at the bottom of a
normed tube
Residual illumination intensity at the bottom of a
normed tube and one shutter
Residual illumination intensity at the bottom of a
normed tube and two shutters
Illumination depth 60 %
Colour rendering index CRI
Colour rendering index R
9
Max. irradiation intensity in the field at a distance
of 1.0 m
Max. irradiation intensity in the field at a distance
of 0.3 m
Light spot size
16
All manuals and user guides at all-guides.com
60,000 Lux
14,000 Lux
103 mm
228 mm
60 mm
103 mm
1180 mm
51 W/m²
330 W/m²
10 cm
Colour temperature (Kelvin)
Heat build-up near the head
Number of LEDs
Working lifespan of the LEDs
Diameter of the light body
Note:
The technical data are subject to certain deviations. Due
to production–related reasons, the actual values can vary
slightly from the above-named values. The values for R
deviate by approx. ± 5%. The colour temperature values can
deviate by approx. ± 200K.
7.2 Electrical data
Power input
Operating voltage
Amperage
7.3 Electrical installation information
The KaWe examination light is exposed to a current peak when it is
switched on and is therefore equipped with a power adapter.
Warning! This light is a Protection Class I device. To avoid the
risk of electric shock, this unit must be connected to a ground-
0 %
ed electrical outlet.
78 %
100%
7.4 Weights
Examination Lamp
0 %
MASTERLIGHT LED 2000
78 %
* with 5-foot stand
95
7.5 Environmental conditions
94
Temperature
Relative humidity
Air pressure
* for higher temperatures please contact us
4000 K
0.5 °C
7
40,000 h
22 cm
can
a
10 W
24 V DC
0,42 A
Weight
6.0 kg*
Operation
Min.
Max.
+10° C
+30° C*
30 %
75 %
700 hPa