Mounting the Warming Unit on an IV Pole
Caution: To prevent tipping, mount the Model 875 warming unit on an IV pole at a height
that ensures stability� We recommend mounting the warming unit no higher than 44 in�
(112 cm) from the floor on an IV pole with a minimum wheelbase radius of 14 in� (35�6 cm)�
Failure to do so may result in IV pole tipping, leading to potential catheter site trauma and
other patient injuries�
1�
Position the IV pole-mount at the desired height on the IV pole�
2�
Turn the clamp knob on the IV pole-mount clockwise to tighten the clamp to the pole�
3�
Snap the warming unit into position by sliding the clip on the back of the warming
unit into the groove of the IV pole-mount� The assembly will click when the clip is fully
seated in the IV pole-mount�
Mounting the Warming Unit on a Bedrail
1�
If the warming unit is not already connected to
the rail-mount, slide the clip on the back of the
warming unit into the groove of the rail-mount�
The assembly will click when the clip is fully
seated in the rail-mount�
2�
Hang the Model 875 warming unit on a bedrail
by the hooks on the rail-mount�
3�
Wrap the safety strap around the bedrail and
connect the end to the fastener on the strap� The
safety strap will prevent the warming unit from
falling if the unit is inadvertently dislodged from the bedrail�
Arizant Healthcare Inc�
Clip
Clamp knob
IV pole-mount
IV pole
Rail-mount
Bedrail
Safety strap
5