Safety System Characteristics
thermostat
overcurrent protection
safety feature
certifications
classifications
Electrical Characteristics
blower motor
power consumption
leakage current
heating element
power cord
device ratings
fuses
temperature controller cord
Arizant Healthcare Inc�
Independent electronic and heater (electromechanical)
Dual input fused line�
Over-temperature protection: color indicator light
illuminates, audible alarm sounds, and heater shuts down
(blower continues to run)�
IEC 60601-1: UL 60601-1; CAN/CSA-C22�2, No� 601�1;
EN 60601-1-2; EN 55011; ASTM F2196-02
Classified under EN 60601-1 Guidelines (and other
national versions of the Guidelines) as Class I, Type BF,
Ordinary equipment, Continuous operation� Not suitable
for use in the presence of flammable anesthetic mixtures
with air or with oxygen or nitrous oxide� Classified by
Underwriters Laboratories Inc� with respect to electric
shock, fire, and mechanical hazards only, in accordance
with UL 60601-1, ASTM F2196-02, and Canadian/CSA
C22�2, No� 601�1� Classified under the Medical Device
Directive (93/42/EEC) as a Class IIb device�
Airflow: 7-13 cfm (3�3-6�1
/
L
S
Peak: 600W Average: 340W
Meets IEC 60601-1 and UL 60601-1 requirements
480W Resistive
15-foot, SJT, 3 cond�, 15A
4�6 m, HAR, 3 cond�, 10A
110-120 VAC, 50/60 Hz, 4�6A
220-240 VAC, 50/60 Hz, 2�8 A
6�0A (120VAC); 4�0A (220-230VAC)
32" from hose collar, 4 cond�, Max� voltage: 5V
)
13