Using the Holder for the Hand-held Temperature Controller
Store the hand-held temperature controller in the controller holder to help prevent cord
clutter around the patient and accidental damage to the warming unit�
1�
Press the holder onto the warming unit hose so it is within the patient's reach�
2�
Slide the temperature controller into the holder� It should fit securely with the dial
facing the patient�
Warming unit
hose
What to Do if the "Over-Temp" Indicator Light
Illuminates and Audible Alarm Sounds
If the "Over-Temp" indicator light illuminates, it means that the warming unit is delivering air at
a higher temperature than the unit is calibrated to deliver� When an over-temperature condition
occurs, the color indicator light illuminates, an audible alarm will sound, and the heater will
automatically turn off (the blower will continue to run)� You should perform the following steps:
1�
Discontinue patient warming�
Disconnect the warming unit hose from the Bair Paws warming gown�
•
Turn the warming unit
•
2�
Unplug the warming unit from the power source and wait 5 minutes�
3�
Reconnect the warming unit to a proper power source�
4�
Turn the warming unit
by turning the controller knob completely clockwise�
5�
Allow the warming unit to run for at least 5 minutes�
If the "Over-Temp" light illuminates again, return the warming unit to Arizant
•
Healthcare for service�
Otherwise, reconnect the warming unit hose to the Bair Paws gown to resume
•
patient warming�